July 13, 2015
100
/100
Combined Rating
100 /100
General Rating
+
General rating
100 /100
1 Ratings
Download
Copy Citation
List of Image Use Citations
Anesthesiology
Image
511 Views
Publication Info
Publication Text
Author(s)
Comments
Publication Info
Received
7/13/2015 5:02:14 PM
Published
7/13/2015 5:12:53 PM
Copyright
(c) 2018 JavMed. All Rights Reserved
License
Author(s)
David Edwards, MdPhD
Conflicts of interest
The author is an editor at JavMed.com
Source of funding
none
Publication Text
Description

The Variable Bypass Vaporizer

The Sevoflurane Vaporizer is an example if a variable bypss vaporizer that delivers vaporized anesthetic gas into an anesthesia machine for inhalation during surgery. Fresh gas flow enters the vaporizer and becomes saturated with anesthetic gas, then exits the other end. A concentration dial at the top allows the user to adjust the output concentration as desired. A bimetallic strip leans to one side as temperature changes occur, automatically adjusting to ensure concentration output is as dialed.

Desflurane Vaporizer

The Desflurane vaporizer is modified from the traditional bypass vaporizer. It has two parallel circuits, each with a resister. In the first circuit fresh gas flows and is regulated by a resister. In the second circuit desflurane is heated in a sump. As the pressure rises a shut off valve opens and allows desflurane to exit through a resister.